v2.1.2 — Hot Fix Release October 2, 2026

Features​

Encapsulated PDF Display Fix​

Hotfix 2.1.2 corrects a defect in which DICOM Encapsulated PDF objects did not display in AdvaView. No new functionality is introduced in this release.

Issue addressed: Opening a DICOM Encapsulated PDF displayed the message "No online PDF viewer installed" in place of the document, and the PDF series thumbnail did not display properly.

Behavior after this hotfix:

  • Encapsulated PDFs now render inside the AdvaView viewport using the browser's built-in PDF viewer, and the "No online PDF viewer installed" message no longer appears.
  • PDF series thumbnails now display correctly in the thumbnail panel.
  • Standard (non-PDF) image series load and display unchanged.
info

PDF rendering in this release depends on the browser's built-in inline PDF viewer. A future release will implement PDF.js-based rendering to convert the PDF into a standard viewable image.


v2.1.1 — Hot Fix Release September 23, 2026

Features​

Platform Not Supported Warning on iOS​

This patch release adds a clear "not supported" warning when AdvaView is opened on iOS devices. Previously, loading the viewer on an iPhone or iPad gave no indication that the platform is unsupported, and the application could hang the device.

Message displayed to users:

Device Not Supported AdvaView is not supported on iOS devices. This applies to all browsers on iOS due to Apple's WebKit requirement. Please access AdvaView from a supported desktop workstation using Google Chrome or Microsoft Edge.

Supported Platforms: AdvaView is supported on desktop workstations using Google Chrome or Microsoft Edge. iOS devices (iPhone and iPad) are not supported.


v2.1.0 - Production Release - Sept 17, 2026

Features​

Study Navigation & Thumbnail Rail​

AdvaView Standard 2.1.0-STD introduces an updated study navigation experience, significant enhancements to Hanging Protocols and multi-monitor setups, clearer loading indicators for advanced 3D/MPR reconstructions, and more intuitive measurement workflows.

  • Upgraded thumbnail rail and study list, now standard, featuring customizable sizing (75%–150%), flexible placement (Top, Left, Bottom), and smooth multi-monitor spanning.
  • Added Quick Drop for drag-and-drop image placement, displaying an instant preview of all available viewports across your displays.
  • Added hover-intent background prefetching for faster series previewing.

Hanging Protocols & Multi-Stage Navigation​

  • Added support for sequential study stages (e.g., current overview followed by prior comparisons) via the new toolbar Stage Navigator.
  • Protocols now scale seamlessly across up to 4 monitors, with automatic mismatch alerts.
  • Added Quick Build preset tools and more flexible study selection criteria for faster creation and management of reading layouts.

MPR & 3D Volume Feedback​

  • MPR viewports now display an in-viewport Rendering Progress (%) indicator during volume loading.
  • Image processing controls (e.g., Maximum Intensity Projection, slab thickness) are now held until volume data is fully loaded, ensuring accurate visualization.

Measurement & Annotation Tools​

  • Added universal cancellation of in-progress measurements via Escape or Right-Click.
  • Added support for the Delete hotkey and standard Enter confirmation dialogs for cleaner keyboard navigation.
  • Enhanced precision and geometry constraints for Cardiothoracic Ratio (CTR), spine labeling, and spline area/perimeter measurements.

Standard Mode Enhancements & Security​

  • Added support for viewing DICOM Structured Reports (SR) in Standard Mode.
  • Enhanced integration security and clean worklist context switching for multi-user reading workstations.

v2.0.2 — Hot Fix Release August 14, 2026

Features​

Software Quality and Security Improvements​

This hotfix release focuses on software quality and security. It strengthens validation of medical imaging data, improves the accuracy of the software component inventory delivered for the product, updates older third-party utilities, and reviews newly identified third-party security findings.

Highlights

  • Improved imaging data validation — Added stronger checks for certain numeric values in medical imaging files to help prevent invalid data from being processed.
  • Improved file handling safeguards — Strengthened controls around how file paths are handled to reduce the risk of unintended file access or writes.
  • More accurate software inventory documentation — Updated the software bill of materials so it better reflects the components included in the production package.
  • Third-party software maintenance — Continued cleanup of outdated or deprecated software packages and review of newly reported third-party vulnerabilities.

v2.0.1 — Hot Fix Release August 6, 2026

Features​

Streamlined DICOM Support​

AdvaView has removed support for RT Structure Set (RTSTRUCT) objects. RTSTRUCT series will no longer appear in the thumbnail rail or be available for selection in the viewer. Studies containing RTSTRUCT data alongside other modalities (e.g., CT, MR) will continue to load and display normally for all supported series types. If a study contained ths SOP class AdvaView woudl report an error to teh end user before loading the rest of the series normally.


v2.0 - Production Release - July 24 2026

Features​

Application Branding Changes​

As part of the release, The application has been completelyrenamed and rebranded to AdvaView with a new color scheme and color palette and new graphic icons, aligned with the corporate look and feel for Adva Health Solutions applications.

Hanging Protocols​

  • Implementation of Hanging Protocol support, including modality layout configuration, study selection criteria with AND/OR logic, display set configuration, and auto-applied protocols.
  • Added window/level preset rules and a rule builder UI for authoring custom display set selection criteria.
  • Added filtering of the HP list by study context (modality) for streamlined protocol selection.
  • Added support for relevant priors configuration within hanging protocol definitions.
  • Added asterisk indicator on the Priors button when prior studies are available.
  • Updated Hanging Protocol selector UI to an icon with an expandable menu.

Multiple Monitor Support​

  • Implementation of Multiple Monitor Support, including dual monitor layouts, floating dialog handling, and fallback messaging for unsupported study types on secondary monitors.
  • Added configuration options to hide thumbnails and toolbar on secondary monitors.
  • Image loading bar replicated across all viewer windows in Multi Monitor mode.

MPR & Crosshairs​

  • MPR and crosshair improvements, including reimplementation of the MPR button and improved crosshair behavior across layout transitions.
  • Added SLAB thickness display showing current value alongside images.
  • Added MNIP (Minimum Intensity Projection) option while in MPR mode.
  • Added Viewport Projection Selection Widget to allow users to select and switch viewport orientation directly from the viewer.

New Tools​

The following diagnostic tools have been added in this release:

  • Overlay Ruler (Scale) — Visual scale overlay displayed on images; toggleable on/off by the user.
  • **Projection Functions ** — Maximum/Minimum/Average Intensity Projection available in MPR mode for enhanced soft-tissue visualization.
  • Augmented Scroll Bar — Visual indicator showing which images in the series have been viewed. Toggleable via hotkey.
  • Grease Pencil — Freeform annotation tool for free-draw markings on any study.
  • Cardiothoracic Ratio (CTR) — Automated cardiothoracic ratio measurement tool for cardiac assessment on chest radiographs.
  • Spine Labeling — Annotation tool for labeling vertebral levels on sagittal, coronal, and axial series.
  • Viewport Projection Selection Widget — Interactive widget enabling users to select and switch viewport orientation directly from within the viewer.

Viewer​

  • Upstream merge to OHIF 3.9 and Cornerstone3D v3.
  • Updated general application color scheme to the new corporate palette as part of the FlexView to AdvaView rebrand.
  • Added AdvaView documentation link to the Gear icon menu.
  • Updated default DICOM metadata overlays for lower-left and lower-right viewport corners.
  • Updated Settings menu labelling for clarity.
  • Updated About screen with revised product logo, links, and Indications for Use statement.

Known Limitations​

  • Not yet recommended for use on mobile devices.

v1.8 — Internal Release Only
Internal Release Only

This version is an internal release and has not been made available to customers.

Rebranding: FlexView → AdvaView

This release marks the complete rebranding of the viewer from FlexView to AdvaView, aligned with the Adva Health Solutions corporate identity. This includes a new application color scheme reflecting the updated corporate palette, updated product logo and links throughout the application, and removal of legacy FlexView AWS Cognito integration from the AdvaView client.

Features​

Viewer​

  • Continued Hanging Protocol development including window/level preset rules, display set selection criteria rule builder UI, filtering of the HP list by study context (modality), and support for relevant priors configuration.
  • Updated general application color scheme to the new corporate palette as part of the FlexView to AdvaView rebrand.
  • Added configuration options to hide thumbnails and toolbar on secondary monitors in Multi Monitor mode.
  • Added an asterisk indicator on the Priors button when prior studies are available.
  • Added AdvaView documentation link to the Gear icon menu.
  • Updated default DICOM metadata overlays for lower-left and lower-right viewport corners.
  • Updated Settings menu labelling for clarity.
  • Updated Hanging Protocol selector UI to an icon with an expandable menu.
  • Segmentation panel disabled pending further review.
  • Image loading bar now replicated across all viewer windows in Multi Monitor mode.

Bug Fixes​

Measurements & Tools​

  • Fixed Livewire measurement showing an empty value in the measurement panel in MPR mode, and Ellipse and Livewire tools not tracking in MPR.
  • Fixed calibration inconsistencies when switching between Stack and Volume modes, and incorrect unit labels persisting after deleting a calibration.
  • Fixed image distortion and artifacts when panning in a non-fullscreen browser window, and images appearing stretched after resizing the browser and dismissing the Capture dialog.

Viewer & Layout​

  • Fixed multiple Radial Menu issues including the MPR icon being broken or missing for new users, inaccessibility for PT modality types, and Alignment and Lock not being greyed out for 3D.
  • Fixed color inconsistencies, shading issues, and prior thumbnail border colors introduced by the rebrand, along with toolbar not respecting modality-specific configuration.
  • Fixed study panel issues including incorrect instance count display, Spine Labels being creatable during viewport load, and reference line markers not being removed in 1x1 layout.

Known Limitations​

  • Saving measurements requires a STOW endpoint which may not be supported by some DICOMWeb datasources.
  • Some datasources (e.g., Amazon HealthImaging) may not support all search queries; search or filter on the study list may return no matches.
  • Not yet recommended for use on mobile devices.

Known Anomalies​

No changes from previous release. Please refer to prior version history for outstanding anomalies.


v1.7 — Internal Release Only
Internal Release Only

This version is an internal release and has not been made available to customers.

Features​

Viewer​

  • Phase 1 implementation of Hanging Protocol support, including modality layout configuration, study selection criteria with AND/OR logic, display set configuration, and auto-applied protocols.
  • MPR and crosshair rewrite, including reimplementation of the MPR button and improved crosshair behavior across layout transitions.
  • Updated About screen with revised product logo, links, and Indications for Use statement.
  • Added SLAB thickness display showing current value alongside images.

Bug Fixes​

Hanging Protocols​

  • Fixed multiple Hanging Protocol creation and editing issues including duplicate name validation, missing confirmation prompts, and incorrect fallback criteria being carried over when copying a protocol.
  • Fixed layout and viewport issues including merged viewports persisting after cancel, non-adjacent viewports disappearing in Multi Monitor mode, and wrong view when entering MPR from an auto-loaded protocol.
  • Fixed HP dropdown not updating after a layout change from an auto-applied protocol, and default modality rules not being recognized.

Measurements & Tools​

  • Fixed measurement orientation not being preserved during navigation after rotate, and missing measurement feedback in MPR view.
  • Fixed multiple Spine Labeling issues including labels moving globally across series, connector line angle errors when moving renamed label groups, crash when restoring MPR layout after labels applied on non-acquisition plane, and inconsistent label hiding in stack mode.

Viewer & Layout​

  • Fixed MIP state loss and crosshair reset when returning from One-Up on a MIP series, and reference lines displaying incorrectly after One-Up transitions when crosshairs have been rotated.
  • Fixed crosshair rotation becoming non-perpendicular after One-Up in MPR, and crosshairs persisting in One-Up view in Multi Monitor mode.
  • Fixed blank or missing viewports when changing layouts, MPR defaulting to 2x2 instead of 1x3, and capture window dismissal resetting projection and crosshairs states.
  • Fixed reset not returning to DICOM window/level on MR modality, and saving measurements via STOW not working.

Known Limitations​

  • Saving measurements requires a STOW endpoint which may not be supported by some DICOMWeb datasources.
  • Some datasources (e.g., Amazon HealthImaging) may not support all search queries; search or filter on the study list may return no matches.
  • Not yet recommended for use on mobile devices.

Known Anomalies​

No changes from previous release. Please refer to prior version history for outstanding anomalies.


v1.6 — Internal Release Only
Internal Release Only

This version is an internal release and has not been made available to customers.

Features​

Viewer​

  • Phase 1 implementation of Multiple Monitor Support, including dual monitor layouts, floating dialog handling, and fallback messaging for unsupported study types on secondary monitors.
  • Added a Viewport Projection Selection Widget to allow users to select and switch viewport orientation directly from the viewer.
  • Renamed the "Zoom In" tool for improved clarity.
  • Added tooltip for the Orientation icon.
  • Added measurements to CSV export.

Bug Fixes​

Measurements & Tools​

  • Resolved tool accuracy failures for Length and Bidirectional measurements (VER-006-006, VER-006-007).
  • Fixed display unit issues in US modality for Ellipse, Bidirectional, and Length tools.
  • Fixed Freehand Length and Freehand Perimeter measurement errors.
  • Fixed Spine Labeling measurements not being tracked on Sagittal and Coronal series in MPR, and labels moving globally across all series regardless of frame of reference.
  • Fixed scale overlay always showing mm and not responding to calibration, including multiple calibration unit and behavior inconsistencies across layouts and modalities.

Viewer & Layout​

  • Resolved multiple Cornerstone regression issues introduced by the upgrade, including stack scroll failures, blinking when opening side panels, incorrect measurement units for Bidirectional and Ellipse tools, and image stack not starting at the first image on load.
  • Fixed MPR series incorrectly carrying over to common layout after One-Up toggle, and series appearing dark after switching from 3D to MPR.
  • Fixed viewer crashes caused by WebGL context loss, RTSTRUCT modality viewing, DICOM Tag tool activation on a blank viewport, and toggling the study panel with the segmentation panel open.
  • Fixed Reset not reverting Window Level and Color LUT, and Projection Function Menu not updating after reset.
  • Fixed overlay retention issues during flip and rotate, and image misalignment after sequential rotate, align lock, and capture operations.

Known Limitations​

  • Saving measurements requires a STOW endpoint which may not be supported by some DICOMWeb datasources.
  • Some datasources (e.g., Amazon HealthImaging) may not support all search queries; search or filter on the study list may return no matches.
  • Not yet recommended for use on mobile devices.

Known Anomalies​

No changes from previous release. Please refer to prior version history for outstanding anomalies.

v1.5 — Internal Release Only
Internal Release Only

This version is an internal release and has not been made available to customers.

Features​

Viewer​

  • Upstream merge to OHIF 3.9 and Cornerstone3D v2.
  • Hidden the US Directional tool pending further validation.
  • Hidden the Hanging Protocol menu item from user-facing settings.

Bug Fixes​

Measurements & Tools​

  • Resolved multiple tool accuracy issues including Bidirectional, Length, Livewire, Spine Labeling, CTR, and Freehand ROI measurements.
  • Fixed measurement tracking failures in the measurement panel for several tools, including Grease Pencil, CTR, and Spine Labeling in standard and MPR modes.
  • Fixed export errors affecting annotations, CSV output, and RT Report generation.
  • Corrected incorrect units of measurement for XC modality type and resolved CTR cardiac shadow miscalculation.
  • Fixed viewer crashes triggered by accessing the measurement panel after use of certain annotation tools.

Viewer & Layout​

  • Fixed multiple layout transition issues including blank viewports when switching between MPR and common layouts, and viewer crashes when switching layouts with active Cine tools.
  • Resolved image orientation and flip issues introduced by the Cornerstone3D 2.0 update, including rotation not being maintained after scrolling.
  • Fixed incorrect display of reference lines across unaligned series, which posed a patient safety risk.
  • Fixed MPR not working in Diagnostic mode and resolved stack image sync with prior not working.
  • Fixed console errors causing undefined Volume ID and passive event listener conflicts that prevented correct image rendering.

Known Limitations​

  • Saving measurements requires a STOW endpoint which may not be supported by some DICOMWeb datasources.
  • Some datasources (e.g., Amazon HealthImaging) may not support all search queries; search or filter on the study list may return no matches.
  • Not yet recommended for use on mobile devices.

Known Anomalies​

No changes from previous release. Please refer to prior version history for outstanding anomalies.


v1.4.2 — November 2025

Features​

This release includes updates to the AdvaView Diagnostic Viewer.

  • Several bug fixes and enhancements to the overall viewer.

Known Limitations​

  • Saving measurements requires a STOW endpoint which may not be supported by some DICOMWeb datasources.
  • Some datasources (e.g., Amazon HealthImaging) may not support all search queries.
  • Not yet recommended for use on mobile devices.

Known Anomalies​

  • Cannot view MPR or PET SUV in Safari v17.1; these features work best in Google Chrome.
  • Issues with segmentation loading times that may cause browsers with limited GPU to hang. Fix expected in v1.4.3.
  • Overlays are not retained correctly when rotate and flip functions are applied.

v1.4.1 — January 2025

Features​

This release includes a number of new tools for the diagnostic version.

  • Added Overlay ruler (scale) which can be toggled on and off.
  • MNIP Minimum Intensity Projection option while in MPR mode.
  • Augmented scroll bar now gives a visual indicator for images in the slice that have been viewed. Can be toggled on or off by hotkey.
  • Pencil tool for freeform annotations on any study.
  • Added Cardiothoracic ratio (CTR) measurement tool.
  • Align and Lock tool allows the user to align the image to either the right or left side of the viewport.
  • Added Spine labeling tool for annotation of the vertebrae of the spine.
  • Added ability to customize viewport labels by user and/or group.
  • Added a color border to visually differentiate comparison studies in the viewport.
  • Added ability to hide viewport labels and annotations with configurable hotkeys.
  • Fixed issue where calibration values might be lost when going in and out of MPR view.

Known Limitations​

  • Cannot view MPR or PET SUV in Safari v17.1; these features work best in Google Chrome.
  • Issues with segmentation loading times that may cause browsers with limited GPU to hang.

Known Anomalies​

  • Cannot view MPR or PET SUV in Safari v17.1; these features work best in Google Chrome.
  • Issues with segmentation loading times that may cause browsers with limited GPU to hang. Fix expected in v1.4.3.
  • Overlays are not retained correctly when rotate and flip functions are applied.

FlexView is Now AdvaView

Adva Health Solutions acquired Radical Imaging's FlexView and rebranded it as AdvaView, now integrated as a viewer module within the AdvaPACS cloud-native imaging platform.


v1.4.0 — May 2024

Features​

A new minor release packed with goodies.

  • New Query Parameters: Added support for flexviewToken, baseUrl, dataSourceToken, and extraHeaders. These enable launching the viewer in a new tab without relying on the postMessage API.
  • New Dropdown Menu including About and User Settings options.
  • Removal of dataSources query parameter, replaced by defaultDatasourceName in viewer configuration.
  • Viewer configuration now includes hotkeys and groupHotkeyPresets properties.
  • Preferences Menu replaced by the new User Settings option.
  • Added new tools: spline, livewire, freehand, ultrasound bidirectional, and icon updates.
  • Added new worklist buttons and tooltips.
  • Added VolumeViewport rotation tool.
  • Added new layout selector in the toolbar which includes MPR.
  • Added new Brush tools and moved SUV peak calculation to the PET SUV mode.
  • Enhanced segmentation panel design for PET SUV mode.
  • Improved the Toolbar to enable reactive state synchronization.
  • Added a viewport action menu that includes colormap per viewport, color bars, 3D presets, and 3D volume rendering configurations.
  • Added a new patient info header component.
  • Optimized the resizing process and maintained the zoom level for viewports.
  • Added new 3D hanging protocols for use in the new layout selector.
  • Updated viewer to evaluate which hanging protocol (HP) to apply without retrieving all series in a study, configurable via minDisplaySetsToRunHP.
  • Several bug fixes and enhancements to the overall viewer.

Known Limitations​

  • Saving measurements requires a STOW endpoint which may not be supported by some DICOMWeb datasources.
  • Some datasources (e.g., Amazon HealthImaging) may not support all search queries.
  • Not yet recommended for use on mobile devices.

v1.3.4— April 15, 2024
note

Platform Version 1.3.4. No changes to Viewer code and therefore no Viewer version change.


v1.3.3— March 22, 2024
note

Platform Version 1.3.3. No changes to Viewer code and therefore no Viewer version change.


v1.3.2— February 18, 2024
note

Platform Version 1.3.2. No changes to Viewer code and therefore no Viewer version change.

Features​

  • Updated About page with new corporate address, UDI, and manufacture date.
  • (Internal) Added Welcome email to be sent 24 hours after customer sign-up.
  • (Internal) Improved alignment of text boxes in Add Datasource screens.

v1.3.0 — January 20, 2024
note

Platform Version 1.3.0. No changes to Viewer code and therefore no Viewer version change.

Features​

  • Added user documentation for essential DICOM metadata fields.
  • Added new warning messaging for unsupported inputs.
  • Improved browser memory detection and user warnings.
  • Resolved rounding error issue in the onscreen display of some calibrated measurements.
  • Updated indications for use to reflect the latest updates to the text.
  • Fixed issue where users could not renew their inactive subscriptions in test environments.

Known Limitations​

  • Saving measurements requires a STOW endpoint which may not be supported by some DICOMWeb datasources.
  • Some datasources (e.g., Amazon HealthImaging) may not support all search queries; search or filter on the study list may return no matches.
  • Not yet recommended for use on mobile devices.

Known Anomalies​

No changes. Please refer to the Viewer version history.


v1.2.10 — November 4, 2024

Features​

No changes were made to the DICOM viewer functionality in this release.

Platform Features​

  • Added a new data source selector dropdown on the Study List page, allowing users to quickly switch between available data sources without navigating to the settings page.
  • Added two new properties to the viewer configuration: hotkeys and groupHotkeyPresets.
  • For customers who embed FlexView and launch it in a standalone tab, added support for the following query parameters: flexviewToken, baseUrl, dataSourceToken, extraHeaders.
  • OpenID Connect (OIDC) property has been moved from a top-level configuration property to being specific to each data source, allowing multiple data sources with OIDC authentication and seamless switching between them.
  • Default DataSource can now be set at the user level instead of the tenant level for more granular access. If a user has no default data source set, it defaults to the first data source defined in the tenant.
  • Updated API: support for dataSources query parameter has been eliminated in favor of defaultDatasourceName set on the viewer configuration, reducing errors such as data source not found or crashes.
  • Added OIDC support for Static Wado data sources.

Viewer Features​

  • Added a new dropdown menu to the FlexView Viewer for customers accessing via iFrame, including:
    • About: shows information about the viewer.
    • User Settings: allows users to change hotkeys or select from available group hotkey presets.
  • Renamed the Preferences menu to "User Settings".

Known Limitations​

  • Saving measurements requires a STOW endpoint which may not be supported by some DICOMWeb datasources.
  • Some datasources (e.g., Amazon HealthImaging) may not support all search queries.
  • Not yet recommended for use on mobile devices.

Known Anomalies​

  • 231 — Stack image sync toolbar is out of sync with actual sync state

    • Description: When displaying a study with series that are auto-image synced, if the user turns "off" the stack sync, it will turn "on" instead. Selecting it a second time will actually turn it "off". When dragging display sets into view that are syncable, the image stack sync will be "on" in the new viewports, even though stack sync is "off".
    • Impact to Users: No impact. Auto-image sync can only be enabled in a hanging protocol, and it is not in use for any hanging protocols in FlexView Diagnostic. Hanging protocols are not configurable by users; thus, the end user cannot create this state.
    • Workaround: This anomaly does not affect users of FlexView Diagnostic. No user workaround is required.
  • 537 — Rounding error in onscreen display of some calibrated measurements

    • Description: There is a slight discrepancy in displayed measurements due to a computational rounding error. This occurs when the software calculates the length of a line in pixels and converts it to millimeters based on the predefined pixel-to-millimeter ratio.
    • Impact: This issue does not occur consistently but may be more noticeable at certain levels of image zoom. The rounding error is minimal, generally within +/- 1% of the actual line length. For example, a calibrated measurement of 10mm might display as 9.99mm.
    • Workaround: No workaround needed; this anomaly will be fixed before the device is available to users.

v1.2.9 — June 13, 2024

Features​

  • Hotfix — Implemented a fallback for TransferSyntaxUID when it is missing in DICOM metadata, allowing images with unknown TransferSyntax to still be displayed.

Known Limitations​

  • Saving measurements requires a STOW endpoint which may not be supported by some DICOMWeb datasources.
  • Some datasources (e.g., Amazon HealthImaging) may not support all search queries.
  • Not yet recommended for use on mobile devices.

Known Anomalies​

  • 231 — Stack image sync toolbar is out of sync with actual sync state

    • Description: When displaying a study with series that are auto-image synced, if the user turns "off" the stack sync, it will turn "on" instead. Selecting it a second time will actually turn it "off". When dragging display sets into view that are syncable, the image stack sync will be "on" in the new viewports, even though stack sync is "off".
    • Impact to Users: No impact. Auto-image sync can only be enabled in a hanging protocol, and it is not in use for any hanging protocols in FlexView Diagnostic. Hanging protocols are not configurable by users; thus, the end user cannot create this state.
    • Workaround: This anomaly does not affect users of FlexView Diagnostic. No user workaround is required.
  • 537 — Rounding error in onscreen display of some calibrated measurements

    • Description: There is a slight discrepancy in displayed measurements due to a computational rounding error. This occurs when the software calculates the length of a line in pixels and converts it to millimeters based on the predefined pixel-to-millimeter ratio.
    • Impact: This issue does not occur consistently but may be more noticeable at certain levels of image zoom. The rounding error is minimal, generally within +/- 1% of the actual line length. For example, a calibrated measurement of 10mm might display as 9.99mm.
    • Workaround: No workaround needed; this anomaly will be fixed before the device is available to users.

v1.2.8 — May 1, 2024

Features​

No changes were made to the DICOM viewer functionality in this release.

Platform Features​

  • Added access to e-label for iFrame customers.
  • Updated process for login and password change for iFrame customers, including a unique URL for partners to sign up.
  • Fixed FlexView logo not showing in the viewer for iFrame customers.
  • Improved the way the viewer iFrame handles messages from the parent application; all messages are now handled by a centralized messaging service.
  • Added a feature allowing resources to be opened by series UID or image UID directly within the iFrame.
  • Published documentation for iFrame customers on setting up DICOM JSON and launching the viewer focused on a specific study.
  • Redesigned the help icon in the lower right corner to be less intrusive.
  • Minimized required permissions for GCP datastore access, enhancing security.
  • Updated the /api/login endpoint to accommodate tenants registered via Auth0, while ensuring compatibility with existing legacy users.
  • Updated About page with new corporate address, UDI, and manufacture date.
  • Added ability to load studies from a location on the desktop.
  • Added a wizard on new account creation to guide users through the setup of their first datasource.

Known Limitations​

  • Saving measurements requires a STOW endpoint which may not be supported by some DICOMWeb datasources.
  • Some datasources (e.g., Amazon HealthImaging) may not support all search queries.
  • Not yet recommended for use on mobile devices.

Known Anomalies​

  • 231 — Stack image sync toolbar is out of sync with actual sync state

    • Description: When displaying a study with series that are auto-image synced, if the user turns "off" the stack sync, it will turn "on" instead. Selecting it a second time will actually turn it "off". When dragging display sets into view that are syncable, the image stack sync will be "on" in the new viewports, even though stack sync is "off".
    • Impact to Users: No impact. Auto-image sync can only be enabled in a hanging protocol, and it is not in use for any hanging protocols in FlexView Diagnostic. Hanging protocols are not configurable by users; thus, the end user cannot create this state.
    • Workaround: This anomaly does not affect users of FlexView Diagnostic. No user workaround is required.
  • 537 — Rounding error in onscreen display of some calibrated measurements

    • Description: There is a slight discrepancy in displayed measurements due to a computational rounding error. This occurs when the software calculates the length of a line in pixels and converts it to millimeters based on the predefined pixel-to-millimeter ratio.
    • Impact: This issue does not occur consistently but may be more noticeable at certain levels of image zoom. The rounding error is minimal, generally within +/- 1% of the actual line length. For example, a calibrated measurement of 10mm might display as 9.99mm.
    • Workaround: No workaround needed; this anomaly will be fixed before the device is available to users.

v1.2.7 — April 22, 2024

Executive Summary​

The FlexView Diagnostic client is built on a base of OHIF and Cornerstone 3D, giving it a substantial set of features, functionalities, and capabilities. This version was released with a Minimal set of diagnostic tools.

This is the first publicly available release of FlexView Diagnostic.

Features​

  • Fixed user initials disappearing in the user menu.
  • Fixed message telling users to sign in when already logged into the viewer; it now correctly says "Go back to viewer".
  • Fixed Azure datasources not displaying the number of instances in a study or series in the study list.
  • Fixed users not being prompted to confirm deletion of a measurement when clicking the trash icon in the measurement panel.
  • Fixed Datepicker component in the study list not being cleared when users clear their filters.
  • Improved warning messages for unsupported input types.

Known Limitations​

  • Saving measurements requires a STOW endpoint which may not be supported by some DICOMWeb datasources.
  • Some datasources (e.g., Amazon HealthImaging) may not support all search queries.
  • Not yet recommended for use on mobile devices.

Known Anomalies​

  • 231 — Stack image sync toolbar is out of sync with actual sync state
    • Description: When displaying a study with series that are auto-image synced, if the user turns "off" the stack sync, it will turn "on" instead. Selecting it a second time will actually turn it "off". When dragging display sets into view that are syncable, the image stack sync will be "on" in the new viewports, even though stack sync is "off".
    • Impact to Users: No impact. Auto-image sync can only be enabled in a hanging protocol, and it is not in use for any hanging protocols in FlexView Diagnostic. Hanging protocols are not configurable by users; thus, the end user cannot create this state.
    • Workaround: This anomaly does not affect users of FlexView Diagnostic. No user workaround is required.
  • 537 — Rounding error in onscreen display of some calibrated measurements
    • Description: There is a slight discrepancy in displayed measurements due to a computational rounding error. This occurs when the software calculates the length of a line in pixels and converts it to millimeters based on the predefined pixel-to-millimeter ratio.
    • Impact: This issue does not occur consistently but may be more noticeable at certain levels of image zoom. The rounding error is minimal, generally within +/- 1% of the actual line length. For example, a calibrated measurement of 10mm might display as 9.99mm.
    • Workaround: No workaround needed; this anomaly will be fixed before the device is available to users.